BTS2048-UV
Bi-technology sensor light meter for high quality UV measurements. Features: Low stray light, 190 nm to 430 nm, short measurement time (electronic shutter), Trigger input and output, entrance optic with diffusor for irradiance and spectral irradiance, etc
Description
UV CCD spectroradiometer vs. broadband CCD spectrometer
The spectral responsivity of conventional CCD detectors usually lies within the 200 nm to 430 nm range. Often, this wide spectral responsivity range of the CCD detector is claimed as the responsivity range of the spectroradiometer. However, this fails to consider the spectral response function of the dispersion grating, which further reduces the detector’s responsivity in the UV spectrum. This results in significant errors in the UV measurement signal, primarily through long-wave stray light. The spectral resolution of broadband spectrometers is often not sufficient to guarantee precise measurements of e.g., narrowband UV LEDs.
CCD spectroradiometers that are specifically designed for UV radiation have a constrained spectral range and allow for a very high grating efficiency in connection with a very high spectral resolution. In addition, optical filters can also be used to significantly reduce stray light.
BTS2048-UV CCD spectroradiometer for UV radiation
The BTS2048-UV meets all the requirements of a high-end UV diode array spectroradiometer and is available at an attractive price despite its cutting-edge technology.
One unique feature of the BiTec sensor is its combination of a back-thinned CCD spectrometer and a Si photodiode that offers high linearity levels enabling extremely fast measurements. The fully linearized 2048 pixel CCD detector with thermoelectric cooling offers a very wide dynamic range thanks to its integration time that ranges from 2 µs to 60 s. This enables precise measurements of UV LEDs in a broad intensity range. The design offers high optical resolution of 0.8 nm over the entire spectral measurement range from 190 nm to 430 nm.. The spectrometer is also equipped with two optical filters for automatic low stray light measurements. Such measurements are necessary for broadband UV lamps and UV LEDs in the presence of other light sources. The very high linearity SiC photodiode within the BiTec detector is used for linearization of the CCD or taken as a reference detector. The radiometric responsivity function of the SiC photodiode enables its use independent of the CCD. The radiometric precision can be auto-corrected using the respective spectral data. The device can therefore be used to perform fast measurements on very weak signals, something that makes the BTS20418-UV perfect for integration in goniometers. Despite its compact dimensions (103 mm x 107 mm x 52 mm – L x W x H), the BTS2048-UV spectroradiometer has a remote-controlled filter wheel with two optical filters and a shutter for dark measurement.
Absolute calibration of the absolute irradiance down to 200 nm!
The many years of experience and its well-equipped DAkkS calibration laboratory (D-K-15047-01-00) enables Gigahertz-Optik to offer traceable calibrations down to 200 nm. This broadens the application range of the BTS2048-UV and UV-C-LEDs. For the shortwave spectral range, Gigahertz-Optik GmbH has implemented a special deuterium lamp-based calibration strategy.
Use in front-end and back-end LED test measurements
The BTS2048-UV is perfectly suited for the testing of UV front-end and back-end LEDs in industrial applications. Its CCD detector integrates an electronic zero setting feature of all pixels before a measurement is triggered (electronic shutter). The electronic shutter and triggering of the measurement can be synchronized with a power supply via a trigger port when the test LED is operated in pulsed current mode. The powerful microprocessor only requires 7 ms to transfer a complete dataset to the system computer via the fast LAN interface.
Direct mounting instead of using a light guide
The BTS2048-UV spectroradiometer has a diffusor window and can therefore be used to measure the UV irradiance, incl. spectrum and peak wavelength, without any additional accessory components. With the diffuser window, the BTS2048-UV can also be mounted directly onto accessories such as integrating spheres, radiance lenses, and goniometers in order to measure the radiant power, radiance, and radiance distribution.
User software and developer software
The standard S-BTS2048 user software has a customizable user interface and offers a large number of display and function modules which can be activated when configuring the BTS2048-UV with the respective accessory components from Gigahertz-Optik GmbH. The S-SDK-BTS2048 developer software is offered for the integration of the BTS2048-UV in the customer’s own software.
Calibration
One essential quality feature of photometric devices is their precise and traceable calibration. The BTS2048-UV is calibrated by Gigahertz-Optik’s calibration laboratory that was accredited by DAkkS (D-K-15047-01-00) for the spectral responsivity and spectral irradiance according to ISO/IEC 17025. The calibration also included the corresponding accessory components. Every device is delivered with its respective calibration certificate.
Specifications
General
Short description
UV optimized TE cooled CCD spectroradiometer with a wide dynamic range for CW and short-term measurement of the irradiance, spectrum, and peak wavelength. Accessories for other parameters.
Main features
Compact device. BiTec detector with back-thinned TE cooled CCD (2048 pixels, 0.8 nm optical resolution, electronic shutter), and SiC photodiode. Optical bandwidth correction (CIE214). Filter wheel with shutter and edge filter. Input lens with diffusor window. Cosine field of view.
Measurement range
Spectral: 3E-5 W/(m²nm) to 3E4 W/(m²nm) @325nm. Responsivity from 190 nm to 430 nm.
Integral: 2E5 W/m² to noise equivalent level by 5E-3 W/m²
typical applications
CCD spectroradiometer for design applications. Module for integration in test systems for front-end and back-end LED testing.
Calibration
Factory calibration. Traceable to international calibration standards
Product
typical applications
Lightmeter for spectral Irradiance, Erythema, etc.
Measured Quantity
Spectral irradiance (W/(m² nm)), irradiance (W/m²), peak wavelength, center wavelength, centroid wavelength, Erythema. Option integrating sphere: in addition spectral radiant power (W/nm) and radiant power (W)
Input optics
Diffusor, cosine corrected field of view (f2 ≤ 3 %)
Filter wheel
4 positions (open, closed, optical filters). Use for remote dark current measurement and stray light reduction.
BiTec
Parallel measurement with diode and array is possible, thereby linearity correction of the array through the diode and online correction of the spectral mismatch of the diode through a*(sz(λ)) respectively F*(sz(λ)).
Calibration uncertainty
Spectral irradiance
(200 – 249) nm: | ± 12 % |
(250 – 299) nm: | ± 7 % |
(300 – 399) nm: | ± 5 % |
(400 – 430) nm: | ± 4 % |
Spectral irradiance responsivity (200 – 430) nm
Measurement modes
Standard measurement mode: 200 nm to 430 nm
Out of Range stray light corrected measurement mode (OoR SLC): 200 nm to 430 nm
Stray light corrected bandpass measurement mode (BP SLC): 300 nm to 386 nm
Spectral Detector
Integration Time
2 μs – 60 s *1
spectral range
(190 – 430) nm
Optical Bandwidth
0.8 nm
Pixel resolution
~0.13 nm/Pixel
Number of pixels
2048
Chip
Highly sensitive back-thinned CCD chip, one stage cooled (1TEC)
ADC
16bit (25 ns instruction cycle time)
Peak wavelength
± 0.05 nm
Band-pass correction
mathematical online band-pass correction is supported
Linearity
completely linearized chip >99.6%
Stray Light
Out of Bound method < 1E-4 *3
Base line noise
5 cts *4
SNR
5000 *5
dynamic range
>9 Magnitudes
spectral responsivity
(3E-4 – 3E4) W/(m²nm) @325nm *6*7
typical measurement time
W/m² of a Halogen lampe from (250 – 400) nm
1 | 4,4 s |
10 | 440 ms |
100 | 44 ms |
Integral Detector
Filter
Mathematical adjustment of the responsivity to a rectangular function from 220 nm to 360 nm (SMCF on-line correction to the radiometric function with the measured spectral data).*
* The spectral responsitivity of the diode does not correspond to a rectangular function (not possible with optical filters). When measuring light sources with a spectrum that deviates from the calibration spectrum of the integral detector (UV LED, peak at 405 nm), the measurement result is corrected using SMCF. The uncertainty of this correction depends on the quality of the measured spectrum (noise) and the size of the correction factor (spectral range).
Measurement time
(0.1 – 6000) ms
Measurement range
seven (7) measurement ranges with transcendent offset correction
Calibration
Irradiance ± 6 % *10
Measurement range
(5E-3 – 2E5) W/m² *11
Loading...
spectral responsivity
f2 (directional response/cosine error)
Miscellaneous
Microprocessor
32bit for device control,16bit for CCD array control, 8bit for photodiode control
Interface
USB V2.0, Ethernet (LAN UDP protocol), RS232, RS485
Data transfer
Standard for 2048 float array values via ethernet 7ms, via USB 2.0 140 ms
Input Interfaces
2x (0 – 25) VDC, 1x optocoupler isolated 5 V / 5 mA
Output Interfaces
2x open collector, max. 25 V, max. 500 mA
Trigger
Trigger input incorporated (different options, rising/falling edge, delayed, etc.)
Software
User software S-BTS2048
Optional software development kit S-SDK-BTS2048 for user software set-ups based on .dll‘s in C, C++,C# or in LabView.
Power Supply
With power supply: DC Input 5V (±10 %) at 700 mA
With USB bus (500mA) *8
Dimensions
103 mm x 107 mm x 52 mm (Length x Width x Height)
Weight
500 g
Mounting
Tripod and M6 screw threads
Front adapter UMPA-1.0-HL for use with integrating sphere port-frame UMPF-1.0-HL
temperature range
Storage: (-10 to 50) °C
Operation: (10 to 30) °C *9
Info
*1 It is recommended to perform a new dark signal measurement for every change in the integration time
*2 typical value, the uncertainty of the dominant wavelength depends on the spectral distribution of the LED
*3 typical value, measured 100 nm left of the peak of a cold white broadband LED with and deep blue LED peak
*4 *5 typical value measured without averaging for a 4ms measurement time and full scale control of the array. Averaging results in quadratic rise of the S/N
i.e. quadratic fall of the base noise e.g. averaging to a factor 100 improves the S/N by a factor 10
*6 Minimum 500/1 S/N. Maximum at full scale control.
*7 Irradiation only allowed for a short time so as to avoid thermal damage
*8 during USB connection, not all functions are available due to the limited current supply e.g. no Ethernet and TEC cooling
*9 Device required for temperature stabilization in approx. 25min. In measurement is performed in the warm-up phase, or if measurements are performed
under varying temperatures, dark signal measurement is required for each measurement
*10 With a(Z) correction by a Deuterium lamp
*11 By a spectral power distribution of a deuterium lamp, maximum radiation only allowed for a short time so as to avoid thermal damage